Setup location, Initial setup procedure – Drucker Diagnostics Model 842VES Centrifuge User Manual

Page 6

Advertising
background image

Setup Location:

1. Unpack the centrifuge and verify that all of the supplied equipment is present.
2. Choose a setup location which meets the following criteria:
a) A bench top clearance height of 16” is required in order to open the lid.
b) The clearance envelope is the space around the centrifuge which is required for

safety. Choose a setup location which will allow for a clearance envelope of at least

24” x 24”, (with the centrifuge at the center). No person or hazardous material shall

be permitted in the clearance envelope during operation. The operator time within

the envelope shall be limited to the time necessary for loading, unloading and

centrifuge operation only.
c) Proper ventilation is necessary to prevent the overheating of samples as well as

premature failure of the centrifuge. Choose an area which will allow unencumbered

air flow.
d) The centrifuge is designed to rest on its four rubber feet. No adjustment is necessary

for leveling the centrifuge, however, the surface should be flat and level.
e)

Be sure the outlet is always within reach as the line cord is the means of

emergency

disconnection!

Initial Setup Procedure:

If any problems are found during the initial setup procedure, refer to the troubleshooting

section. For further assistance, contact The Drucker Company at

814-342-6205 or

814-342-6205.

1. Plug the female end of the line cord into the rear of the centrifuge. Plug the male end into

an approved electrical outlet. For electrical safety, the unit must always be properly

grounded.
2. For operator safety, the locking system is always active; requiring power and direction

from the user to disengage it (the lid also automatically unlocks at the end of a run when

it is safe to do so). To unlock the lid (in order to access the rotor chamber) press the

‘OPEN / STOP’ button on the control panel. The ‘UNLOCKED’ indicator light should

illuminate. If it does not, refer to the section on troubleshooting. The lid will remain

unlocked for 15 seconds after pressing the ‘OPEN / STOP’ button.
3. Turn the latch counterclockwise and open the lid.
4. Spin the rotor by hand; check for free and level rotation. If the rotor does not spin freely,

refer to the section on troubleshooting.
5. Place the test tube holders inside the rotor and verify that they are seated properly.
6. Close the lid. Rotate the lid knob clockwise to its complete stop position. The

’LATCHED’ indicator light should be illuminated. If it is not, make sure that the lid is

latched properly. The centrifuge will not run unless the lid is latched properly and the

’LATCH’ indicator light is illuminated.
7. Use the up arrow button next to the RPM display to set the speed to 4000 RPM.
8. Initiate a test run by pressing the ‘START’ button.
9. The ‘RUNNING’ indicator light will illuminate.
10. The test tube carriers will slide up into the horizontal position (horizontal rotor only) and

the unit will accelerate to the current set speed.
11. Listen to the sound of the centrifuge. A smooth whirring sound should be heard. If there

are any loud or unusual sounds, stop the centrifuge by pressing the ’OPEN / STOP’ button

immediately and refer to the section on troubleshooting.
12. Press the ‘OPEN / STOP’ button to terminate the test run. The rotor will decelerate to a

complete stop and the lid will then unlock automatically for sixty (60) seconds.
13. Take time now to familiarize yourself with the various additional user settings. Turn to

page 11.
14. The centrifuge is now ready for operation.

page 6

Advertising