Iris Sample Processing StatSpin® Express 3 Primary Tube Centrifuge User Manual

Page 13

Advertising
background image

9

Troubleshooting

No power light

Is the line cord completely plugged

into the wall?

No→

Plug line cord into the wall.

Yes

Is the wall outlet working properly?

→ No→

Try another wall outlet.

Yes

Is the fuse blown?

→ Yes→

Replace fuse.

No

Contact Iris Sample Processing or

an Authorized Repair Facility.

Centrifuge will not spin, or shuts

off prematurely

Is the cover closed and latched?

No →

Completely close the cover.

Yes

Is the error light flashing?

Yes →

Identify source of error from flashing
indicator lights, correct the error condition,
and push the Stop/Open button to clear the
error.

No

Contact Iris Sample Processing or

an Authorized Repair Facility.

Centrifuge will not open at end

of cycle

Is the bowl gasket completely

seated and sealing the bowl?

NO→

Press the bowl gasket all the way down into
the housing slot.

YES

Contact Iris Sample Processing or

an Authorized Repair Facility.

Advertising