Operator Controls

Start

The Start button initiates the selected cycle.

Stop/Open

The Stop/Open button interrupts the cycle and
releases the cover when the rotor has stopped.

2 Minute Cycle

The 2 min button selects a cycle of two minutes at
7200 rpm.

3 Minute Cycle

The 3 min button selects a cycle of three minutes at
7200 rpm

.

5 Minute Cycle

The 5 min button selects a cycle of five minutes at
5600 rpm.

Error/Service

The red indicator with an adjacent wrench symbol
flashes to signal an error condition or remains
continuously illuminated when service is needed.

Error Indicators

Express 3 error conditions are signaled by a flashing red Error/Service indicator on the control panel.


The cause of the error condition can be identified by the combination of green cycle time indicators
that are also flashing.
















Error/Service

light flashing

Rotor failed to

reach full speed

in 30 seconds.

Cover not

latched

Rotor

over-speed

Rotor

out-of-balance

Unable to

maintain rotor

speed

Locked rotor

2 min

3 min

5 min