How to use this manual – Iris Sample Processing StatSpin® CritSpin Microhematocrit System User Manual

Page 4

Advertising
background image

ii

How to use this manual

This manual along with information contained on product labels and in package inserts should
provide you with all the information you need to operate and maintain the CritSpin.

Notes appear in italics to highlight information. When the information requires special attention, a
caution symbol appears next to the italicized text.

Please pay close attention to the instructions that accompany the notes and symbols as well as the
standard laboratory practices outlined by your facility and local regulatory agencies. The table below
lists all the CAUTIONS/WARNINGS for the CritSpin.

WARNING - North American Installation: Only use the power supply included
with the unit. Use of other power supplies or transformers will damage the CritSpin
electronics and void the warranty.

WARNING - Outside of North America: Do Not Use the Line Cord Supplied. Use
power cord for at least 1.0 Amp or more with an IEC320/CEE22 female connector
and male connector suitable for the power outlet to be used.

CAUTION - The cover interlock bypass is designed for emergency use only. If the
equipment is not used properly, safety may be impaired.

CAUTION – Do not leave any rotor on the rotor-holder when the StatSpin is not in
use for an extended period of time. Doing so may compress the O-Ring and
decrease its ability to hold the rotor.

CAUTION – Failure to properly install the rotor may result in damage to the
centrifuge and will void the warranty.

CAUTION – If rotor is left in place between runs, be certain to bottom the rotor on
the rotor-holder before spinning another sample. Failure to properly seat the rotor
each time may result in the rotor becoming loose during centrifugation.

IMPORTANT- The rotor (RH12) has a finite lifespan that is dependant on usage.
Cracks or swelling that develop indicate the rotor should be replaced immediately.

CAUTION

Unplug the CritSpin from the wall outlet before performing

maintenance.


WARNING – Do not expose CritSpin and its rotor to strong or concentrated acids,
bases, esters, aromatic or halogenated hydrocarbons, ketones or strong oxidizing
agents.

CAUTION – DO NOT spray the bowl or outer surfaces with detergent or bleach.
Excess liquid will harm the electronics and subsequent problems may not be
covered under warranty.

CAUTION - Universal Precautions should be followed on all specimens, regardless
of whether a specimen is known to contain an infectious agent. (See references)

Please use the system as intended. Improper use of the CritSpin Centrifuge and/or its accessories
may cause damage to the system, inaccurate results, or potentially nullify warranties.

The CritSpin Centrifuge, and its associated components are covered under one or more of the
following US patents nos.: 4,846,974 & 4,981,585 & 5,566,919