Iris Sample Processing StatSpin® CritSpin Microhematocrit System User Manual

Page 7

Advertising
background image

3

Error Indicators

Error codes for the CritSpin are specified by the Error/Service LED flashing or continuously
illuminated.

Error/Service =










Error/Service light will continuously illuminate when the centrifuge has achieved a total cycle count of
18,000, which is the useful life of the drive system. Drive mechanism needs replacing. Contact an
authorized service center.

Error/Service light will flash when one of the following conditions occur:

1. System Error – Restart the centrifuge.
2. Motor Failure – Drive mechanism requires replacement.
3. Rotor Improperly Seated – Re-install rotor.

Error/Service and 120 s lights will flash when the cover is open during spin.

Accessories

Product No.

Description

RH12

Hematocrit rotor, 12-position

HR4C

Circular Hematocrit Reader

HR05

Card-style Hematocrit Reader

HT9H-10

Hematocrit Tubes, 40mm, glass, heparin

HT9U-10

Hematocrit Tubes, 40mm, glass, untreated

HP8U-10

Hematocrit Tubes, 40mm, plastic, heparin

HP8U-10

Hematocrit Tubes, 40mm, plastic, untreated

HS24-10

Clay Sealant Pad

CSD1

Digital Reader (included with CS22 Product Number)

A

B

C

120S

A Error/Service light
continuously
illuminated

B Error/Service
light flashing

120S

C Error/Service and
120 s flashing

Advertising