VerifyNow System User Manual |

USA

12.

Close the cover to the test port (Figure 6-11).

Figure 6‑11 Close Cover

13.

The instrument automatically draws the sample from the vacuum collection tube into the test device.

It then heats the blood to 37°C for a period of time specific to each test, and proceeds with the

analysis. During this time the screen will flash indicating that the sample is processing. Do not open

the test port cover until the test is complete. A calculator will display when the test is near completion

(Figure 6-12).

Figure 6‑12 Test Progress

VerifyNow P2Y12

26-Feb-2012

123

14:57:00

KEEP COVER CLOSED.

999

VerifyNow P2Y12

26-Feb-2012

123

14:57:00

KEEP COVER CLOSED.

VerifyNow P2Y12

26-Feb-2012

123

14:57:00

CAUTION:

The sample is pressurized during parts of the test. Do

not remove the sample tube or test device during the test. To abort,

press the Back

key. Wait for the prompt before removing the test

device along with the tube.

14.

It typically takes 3 to 5 minutes for the test to complete. When the results display, record or print

the results, if your instrument has an attached printer.

15.

Open the cover. Remove the test device and tube together in one piece by grasping the test device

finger grip and pulling straight up (Figure 6-13). Do not separate the tube from the test device. Close

the cover. Discard the test device and tube in a biohazard waste in accordance with institutional

guidelines.