Main te nance 11, How to use the main te nance sec tion 11, How to re turn ma te rial to deltech 11 – Deltech Fitness PYRAMID 8000 User Manual

Page 17: Re place ment parts 11, Pre ven tive main te nance sched ule 11, Maintenance, How to use the maintenance section, How to return material to deltech, Replacement parts, Preventive maintenance schedule

Advertising
background image

6. Lift the vessel out from the top. Use the boom and

hoist for models 8DM100 and 8DM150; vessels

for smaller models can be lifted out by hand.

7. All vessels may be discarded, except BA17 and

AC17. These vessels can be returned to the factory

for credit. See page 11 for return procedure.

8. Lower a new vessel into place from the top. Use

the boom and hoist for models 8DM100 and

8DM150; vessels for smaller models can be low-

ered by hand.

9. Close the hinged support clamps to secure the new

vessel to its support bracket.

10. Connect the unions on the top and bottom of the

vessel.

11. Replace all cabinet panels.

12. Record the date of change on the spare label sup-

plied with the replacement vessel; stick the label to

the top of the purifier.

WARNING

To restart the purifier, follow the procedure

described in the Start-Up section of this man-

ual.

Boom Assembly

A boom and hoist assembly are provided with Models

8DM100 and 8DM150. Use the boom and hoist to re-

move and replace the activated carbon filter and cata-

lyst vessel according to the Preventive Maintenance

Schedule in this manual. The boom is stored inside the

cabinet in the back left corner. The hoist is strapped to

the bottom of the purifier cabinet frame.

To use the boom and hoist:

1. Lift the boom assembly out from the top and set it

on one of the bases provided on two sides of the

purifier (each base looks like a closed-off black

pipe sticking up near the top of the purifier).

2. Attach the hoist to the boom hook.

3. Swing the boom hook over the vessel.

4. Attach the hoist to the handles provided on the

vessel.

5. Disconnect the unions on the top and bottom of

the vessel.

6. Using the hoist, lift the vessel out of the purifier

cabinet.

7. Swing the boom to the side and use the hoist to

lower the vessel to the floor.

8. Attach a new vessel to the hoist; using the boom

assembly, lift the vessel above the cabinet and

lower it into place.

9. Store the boom assembly inside the purifier cabi-

net.

Coalescing Filter Element Replacement

To replace the element:

1. Shut down the purifier according to the Shutdown

procedure.

2. Disconnect electrical power to the automatic drain

valves.

3. Remove cabinet panels to expose filter.

4. Carefully remove black foam insulation from filter

bowl and set aside.

5. Mark center drain line and outer drain line for re-

connection, then disconnect the drain lines from

the filter bowl.

6. Remove the threaded coalescing filter bowl from

the top casting, using a strap wrench if necessary.

7. Remove the saturated element from the bowl.

8. Clean the bowl with soap and water. Do not use

solvents.

9. Insert a new element in the bowl.

10. Reattach the bowl to the top casting.

11. Reattach the center and outer drain lines.

12. Replace the black foam insulation and secure it

with tape.

13. Reconnect electrical power to the automatic drain

valves.

14. Replace all cabinet panels.

Pyramid 8000

®

Series Purifiers (Bulletin 289)

13

Advertising