3.2.3 Logging process data

Figure 3.5

Ongoing data logging

You can log both process data and optical image informa-

tion.

To log process data, choose

MEASUREMENT

in the main

screen, then

DATA LOGGING

. Before starting the data logging,

set the

INTERVAL

by touching the grey area. Default value

for the interval is 10 seconds. Start logging by touching the

START

button.

The data is saved into a text file in the FC-11’s internal mem-

ory, or on a USB stick, if available. If a USB stick is con-

nected, data will automatically be saved on it.

Important:

Do not disconnect an external memory (USB

stick) while logging or you risk losing the log.

Note: If data logging is in process some functions are dis-

abled, e.g. the

TOOLS

menu.

See section 2.5.1 for information on how to access the

logged data.