Setting pulse output parameters, Using the output multiplexer, Setting system internal rate parameters – Quantum Composers 9530 Series User Manual

Page 12

Advertising
background image

Page 12

9530/8530 Manual Version 4.8


Setting Pulse Output Parameters

There are two basic types of outputs available on the 9530:

a) TTL/CMOS

compatible outputs.

b)

Adjustable amplitude outputs.

Out:

Selects between TTL/CMOS mode and Adjustable
mode when both are available on a single output.

Pol:

Sets the voltage polarity of the pulse, active high or

active low.

*Note: All outputs are positive voltage as negative
voltages are not supported

Ampl:

In Adjustable mode the unloaded output voltage is set.
The actual output voltage will depend on the load
impedance. For example: If the load is 50Ω the output
will be 50% of the stated voltage.

Using the Output Multiplexer

Each output channel includes a multiplexer which allows routing of any or all of the timer outputs to
the physical output. This allows double pulses and other complex pulse trains to be generated.

-HGFEDCBA-

MUX:

-00000101-

The multiplexer is represented by an

“n” bit binary number as shown above. “N” is the number of

channels. Each bit represents a channel timer, which is enabled by setting the bit to one. In the
above example, timers A and C are combined on the current output.

Setting System Internal Rate Parameters

The internal T

0

period controls the fundamental output frequency of the system. Each channel may

be operated at submultiples of the fundamental frequency by using the channel duty cycle mode.

Source:

Sets the reference source for the internal

T

0

period.

Per:

Sets the internal

T

0

period.

To set the system Internal Rate press the Menu key, then use the arrow keys

to specify the

T

0

period.

Advertising