Monthly maintenance, Checking the solution temperatures, Monthly maintenance -22 – Kodak RP 30 User Manual

Page 214

Advertising
background image

Operator’s Guide

Monthly Maintenance

Checking the Solution Temperatures

1. Touch:

Settings

Machine settings

PP temperature

2. Set the reference temperatures for the solutions and the dryer, if necessary:

DEV

40 °C ± 0.3

BX

38 °C ± 2

STB

37 °C ± 2

Dryer

65 °C

3. Calibrate the temperature sensors:

a. Touch

Readjust xxx.

A window with a numeric keypad opens. The temperature of the selected
solution is indicated in the first line of the input field.

b. Remove the chemical filter.

Using a calibrated thermometer, measure the current temperature of the
processing solution through the free opening of the filter holder.

c. Enter the current temperature value via the numeric keypad.

The sensor will be calibrated after pressing OK.

NOTE: The Reset reference and Reset calibration buttons are used to enter the

standard values of the database.

8-22

KODAK PROFESSIONAL RP 30 Laser Printer