Reference and Power Sources

The NI USB-6008/6009 creates an external reference and supplies a power
source. All voltages are relative to COM unless otherwise noted.

+2.5 External References

The NI USB-6008/6009 creates a high-purity reference voltage supply for
the ADC using a multi-state regulator, amplifier, and filter circuit. The
resulting +2.5 V reference voltage can be used as a signal for self test.

+5 V Power Source

The NI USB-6008/6009 supplies a 5 V, 200 mA output. This source can be
used to power external components.

Note

While the device is in USB suspend, the output is disabled.

Specifications

The following specifications are typical at 25 °C, unless otherwise noted.

Analog Input

Converter type........................................ Successive approximation

Analog inputs ......................................... 8 single-ended, 4 differential,

software selectable

Input resolution

NI USB-6008 .................................. 12 bits differential,

11 bits single-ended

NI USB-6009 .................................. 14 bits differential,

13 bits single-ended

Max sampling rate (aggregate)

1

NI USB-6008 .................................. 10 kS/s

NI USB-6009 .................................. 48 kS/s

AI FIFO.................................................. 512 bytes

Timing resolution ................................... 41.67 ns (24 MHz timebase)