Safety, Safety voltages, Safety standards – National Instruments NI USB-6009 User Manual
Page 30: Hazardous locations, Environmental
NI USB-6008/6009 User Guide and Specifications
30
ni.com
Safety
If you need to clean the module, wipe it with a dry towel.
Safety Voltages
Connect only voltages that are within these limits.
Channel-to-GND ....................................±30 V max,
Measurement Category I
Measurement Category I is for measurements performed on circuits not
directly connected to the electrical distribution system referred to as
MAINS voltage. MAINS is a hazardous live electrical supply system that
powers equipment. This category is for measurements of voltages from
specially protected secondary circuits. Such voltage measurements include
signal levels, special equipment, limited-energy parts of equipment,
circuits powered by regulated low-voltage sources, and electronics.
Caution
Do not use this module for connection to signals or for measurements within
Measurement Categories II, III, or IV.
Safety Standards
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or visit
ni.com/certification
, search by model number or product line, and click the
appropriate link in the Certification column.
Hazardous Locations
The NI USB-6008/6009 are not certified for use in hazardous locations.
Environmental
The NI USB-6008/6009 device is intended for indoor use only.
Operating temperature
(IEC 60068-2-1 and IEC 60068-2-2)......0 to 55 °C
Operating humidity
(IEC 60068-2-56) ...................................5 to 95% RH, noncondensing