Table 17. ni 9205/9206 scan list format – National Instruments Deterministic Ethernet Expansion Chassis NI 9144 User Manual

Page 30

Advertising
background image

NI 9144 User Guide and Specifications

30

ni.com

Table 17 contains the default scan list.

NI 9205/9206 Calibration Data

The NI 9205 uses a quadratic formula for conversion from 16-bit raw data
to calibrated data.

The NI 9205 EEPROM provides overall polynominal values a3–a0 along
with gain and offset values for each voltage range, to be applied when
converting 16-bit raw data to calibrated data.

1.

Convert the 32-bit hex values to 64-bit floating point format for use in
the calibration formula.

2.

Select the 32-bit gain value for a particular range.

Table 17. NI 9205/9206 Scan List Format

Index

Sub

Type

Value

Sub

Value

0x2001

0

ARR:U32

33

1

32

2

0x00002A38

18

0x00103238

3

0x00012B38

19

0x00113338

4

0x00022C38

20

0x00123438

5

0x00032D38

21

0x00133538

6

0x00042E38

22

0x00143638

7

0x00052F38

23

0x00153738

8

0x0006283C

24

0x0016303C

9

0x0007293C

25

0x0017313C

10

0x00082A3C

26

0x0018323C

11

0x00092B3C

27

0x0019333C

12

0x000A2C3C

28

0x001A343C

13

0x000B2D3C

29

0x001B353C

14

0x000C2E3C

30

0x001C363C

15

0x000D2F3C

31

0x001D373C

16

0x000E3038

32

0x001E2838

17

0x000F3138

33

0x001F2938