System Specifications

English

Guidance and Manufacturer’s Declaration - Electromagnetic

Emissions and Immunity

Guidance and manufacturer’s declaration – electromagnetic emissions

The VPAP is intended for use in the electromagnetic environment specified below. The customer or

the user of the VPAP should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment -
guidance

RF emissions CISPR11

Group 1

The VPAP uses RF energy only for its

internal function. Therefore, its RF

emissions are very low and are not

likely to cause any interference in

nearby electronic equipment.

RF emissions CISPR 11

Class B

The VPAP is suitable for use in all

establishments, including domestic

establishments and those directly

connected to the public low-voltage

network that supplies buildings used

for domestic purposes.

Harmonic Emissions

IEC 61000-3-2

Class A

Voltage Fluctuations/Flicker

Emissions IEC 61000-3-3

Complies

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed

and put into service according to EMC information provided in this document.

Warnings: The VPAP should not be used adjacent to or stacked with other equipment.

If adjacent or stacked use is necessary, the VPAP should be observed to verify normal operation in

the configuration in which it will be used.

The use of accessories (eg humidifiers) other than those specified in this manual is not

recommended. They may result in increased emissions or decreased immunity of the VPAP.