VEGA VEGASON 71 - 1 … 75 - 1 User Manual

Page 36

Advertising
background image

Supplementary programmings

36

71 - 1 … 75 - 1

Simulation

MODE 0 - 94

These modes enable the outputs for each channel
to be simulated over the whole range of the
distance in m, to test the function of the connected
instrumentation.

The simulation influences the selected channel and
therefore the outputs coordinated to this channel,
i.e. the
-

relay outputs

-

current outputs and

-

the indication (display 2)

mode

simulation range

no.

72 / 73 - 1

74 - 1

75 - 1

94

0 … 11 m

0 … 22 m

0 … 33 m

Operation

The plus button increases,
the minus button reduces
the outputs.

The simulation operates first
of all 0,05 m steps and is
accelerated to 0,25 m steps
after approx. 10 secs.

Note

The active simulation is recognized by
-

selection of mode 94

-

flashing indication of all figures, e.g. .....................................................

15.25

m.

Note
If the simulation is active the measured values from
the sensor are NOT transmitted.

Mode 0 - 94 must therefore be immediately quit
after the simulation is carried out.

The timer reset (after 60 minutes) should only be
used if absolutely necessary.

After having finished the simulation all outputs are
up-dated with the valid measured value.

Simulation example ...............................................

• select mode 0 - 94 in MODEFIELD
• activate simulation with button MOD the actual

measurement flashes in PARAMETER FIELD

• modify the outputs with button Plus or Minus

• After having finished the test procedures quit the

simulation with MOD- or STO-button.

relay output

measured value

sensor

current output

Simulation:

indication (in m)

15.25

0 - 94