Atec Agilent-E4980A-AL User Manual
Page 33
33
When DC bias is ON, the following time is added:
Table 57. Additional time when DC bias is ON [ms]
Test frequency
20 Hz
100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
2 MHz
30 30 10 13 2
0.5 0.5
When the number of averaging increases, the measurement time is given as
Equation 17.
MeasTime + (Ave – 1) × AveTime
MeasTime
Measurement time calculated based on Table 53 and Table 54
Ave
Number of averaging
AveTime
Refer to Table 56
Table 58. Additional time per averaging [ms]
Measurement
time mode
Test frequency
20 Hz
100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
2 MHz
SHORT 51 11 2.4
2.3 2.3 2.2 2.2
MED 110
81
88
87
85
84 84
LONG 210
210
220
220
220
210
210
Table 59. Measurement time when Vdc-Idc is selected [ms]
Test frequency
Measurement time mode 20 Hz 100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
2 MHz
SHORT 210
46
14
14
14
14
14
MED 210
170
170
170
170
170
170
LONG
410
410 410
410
410 410 410
Add the same measurement time per 1 additional average
Additional Measurement time when the Vdc and Idc monitor function is ON.
Add SHORT mode of Table 59. When using only Vdc or Idc, add a half of SHORT mode of
Table 59.
Table 60. Measurement time when Ls-Rdc or Lp-Rdc is selected [ms]
Test frequency
Measurement time mode 20 Hz 100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
2 MHz
SHORT
910
230
43
24
22
22
22
MED
1100
450
300
280
270
270
270
LONG
1400
820
700
670
660
650
650
Add the three times of measurement time per 1 additional average number