Atec Agilent-E4980A-AL User Manual

Page 33

Advertising
background image

33

When DC bias is ON, the following time is added:

Table 57. Additional time when DC bias is ON [ms]

Test frequency
20 Hz

100 Hz

1 kHz

10 kHz

100 kHz

1 MHz

2 MHz

30 30 10 13 2

0.5 0.5

When the number of averaging increases, the measurement time is given as

Equation 17.

MeasTime + (Ave – 1) × AveTime

MeasTime

Measurement time calculated based on Table 53 and Table 54

Ave

Number of averaging

AveTime

Refer to Table 56

Table 58. Additional time per averaging [ms]

Measurement
time mode

Test frequency

20 Hz

100 Hz

1 kHz

10 kHz

100 kHz

1 MHz

2 MHz

SHORT 51 11 2.4

2.3 2.3 2.2 2.2

MED 110

81

88

87

85

84 84

LONG 210

210

220

220

220

210

210

Table 59. Measurement time when Vdc-Idc is selected [ms]

Test frequency

Measurement time mode 20 Hz 100 Hz

1 kHz

10 kHz

100 kHz

1 MHz

2 MHz

SHORT 210

46

14

14

14

14

14

MED 210

170

170

170

170

170

170

LONG

410

410 410

410

410 410 410

Add the same measurement time per 1 additional average

Additional Measurement time when the Vdc and Idc monitor function is ON.
Add SHORT mode of Table 59. When using only Vdc or Idc, add a half of SHORT mode of
Table 59.

Table 60. Measurement time when Ls-Rdc or Lp-Rdc is selected [ms]

Test frequency

Measurement time mode 20 Hz 100 Hz

1 kHz

10 kHz

100 kHz

1 MHz

2 MHz

SHORT

910

230

43

24

22

22

22

MED

1100

450

300

280

270

270

270

LONG

1400

820

700

670

660

650

650

Add the three times of measurement time per 1 additional average number