Absolute measurement accuracy – Atec Agilent-E4980A-AL User Manual
Page 7
7
Absolute measurement accuracy
The following equations are used to calculate absolute accuracy.
Absolute accuracy Aa of |Z|, |Y|, L, C, R, X, G, B (L, C, X, and B accuracies apply
when Dx ≤ 0.1, R and G accuracies apply when Qx ≤ 0.1 )
Equation 1.
Aa = Ae + Acal
Aa
Absolute accuracy (% of reading value)
Ae
Relative accuracy (% of reading value)
Acal Calibration
accuracy
(%)
where G accuracy is applied only to G-B measurements.
D accuracy (when Dx ≤ 0.1)
Equation 2.
De + θcal
Dx
Measured D value
De
Relative accuracy of D
θcal
Calibration accuracy of θ (radian)
Q accuracy (When Qx × Da < 1)
Equation 3.
±
(Qx
2
× Da)
(1 Qx × Da)
Qx Measured
Q
value
Da
Absolute accuracy of D
θ accuracy
Equation 4.
θe + θcal
θe
Relative accuracy of θ (degree)
θcal
Calibration accuracy of θ (degree)
±