42
waveform, a message to save the
Arbitrary wave will be displayed. Select
YES or NO to save the new waveform.
3.6.7 Arbitrary EDIT Menu
Enters data for creating arbitrary waveforms.
You can enter data one point at a time, as a
value at an address, draw a line from one
point (a value at an address) to another point,
create a predefined waveform, or combine
these to create complex waveforms. The valid
data values range is -2047 to 2047. The valid
waveform memory addresses range from 1 to
1,000.
The data value governs the output amplitude
of that point of the waveform, scaled to the
instrument output amplitude. Therefore, a
value of 2047 corresponds to positive peak
amplitude, 0 corresponds to the waveform
offset, and -2047 corresponds to the negative
peak amplitude
.