B&K Precision 4045B - Manual User Manual

Page 87

Advertising
background image

87

AMPL 2.5V

AMPL MAX

Query

Syntax: AMPLitude? <ws>MINimum|MAXimum]

Examples: AMPL?

AMPL? MAX

Response: NR2

Considerations:

1) The MAXimum amplitude is dependent on the offset.

2) The MAX and MIN arguments should not be used in a

program message containing an OFFSet command, since

these values are evaluated during parsing, based on the

current value of the offset.

OFFSet <offset>

The offset command is used to set the DC offset of the

output waveform. Note that the amplitude and the offset

are limited by the relation:

Peak Amplitude + |Offset| ≤ 5V

Arguments

Type: Numeric

Units: V, mV

Range: 10mV to 5V

Rounding: To 10mV

Command Type: Setting or Query

Setting

Syntax: OFFSet<ws><offset>[units]

Advertising