Installation
Figure 12 SS7 sc-HST plumbing diagram
1
Optional items
14 Sample unit
2
Bubble trap
15 sc100
3
3-way ball valve (Auto Flush Kit)
16 Customer supplied power on/off switch box (NEMA 4X)
required for agency compliance
4
Cooling water to drain
17 Power in for sc100
5
Cooling water out
18 ¾-in. NPT adapter (supplied)
6
Flow control valve
19 Drain Trap (Customer-supplied)
7
Sample in
20 1-in. NPT adapter (supplied)
8
Sample bypass during flush cycle
21 To drain
9
Cooling water in
22 ¼-in. air purge fitting (50 SCFH instrument air max)
10 Heat exchanger
23 Ball valve (supplied)
11 Electrical box connection
24 ¾-in. NPT nipple (supplied)
12 127 mm (5 in.) minimum
25 Customer supplied hose to drain
13 Customer supplied air for flow multiplier
A
Sample during normal operation
D
Cooling water during auto flush
B
Sample bypass during auto flush
E
Electrical
C
Cooling water in normal operation
F
Drain