Appendix A Modbus register
Tag Name
Register #
Data Type
Length
R/W
Description
TURB
40001
Float
2
R
Measured turbidity value
TURB INT
40003
Unsigned Integer
1
R
Integer turbidity value
TURB INT X 100
40004
Unsigned Integer
1
R
Integer turbidity * 100
SENSOR NAME
40005
String
6
R/W
Sensor name or location
BUBBLE REJECT
40011
1
1
In order to write to these tags, write 46478 to register 49938
Unsigned Integer
1
R/W
Bubble reject status (0=OFF; 1=ON)
SIGNAL AVG
40012
1
Unsigned Integer
1
R/W
Signal average (0=1; 1=6sec;
2=30sec;3=60sec;4=90sec)
DATALOG INTRVL
40013
1
Unsigned Integer
1
R/W
Datalog interval
(0=5sec;1=30sec;2=1min;3=2min;4=5min;6=10
min;7=15min;8=30min;9=60min, 10=4hr)
RESOLUTION
40014
Unsigned Integer
1
R/W
Maximum number of decimal places (0=xxxxx,
1=xxxx.x, 2=xxx.xx)
P/F CRITERIA
40015
1
Unsigned Integer
1
R/W
Pass / Fail criteria for verification. (1 to 10
percent)
TURB UNIT
40016
Unsigned Integer
1
R/W
Turb units (0=mg/L:7=NTU;42=FTU)
SERVICE MODE
40018
1
Unsigned Integer
1
R/W
Used to determine if the instrument is in the
service mode (0 = disabled; 1= enabled)
SERIAL NUMBER
40021
String
6
R
Instrument serial number
CODE VERSION
40027
Float
2
R
Software version
DD Firmware
40029
Unsigned Integer
1
R
Device driver firmware version
DD CONTENT
40030
Unsigned Integer
1
R
Device driver content version
HW VERSION
40031
Unsigned Integer
1
R
Hardware version of pc board
TEMP
40032
Float
2
R
Temperature measurement in Celsius
DARK
40034
Unsigned Integer
2
R
Dark turbidity A/D counts.
RAW TURB
40036
Float
2
R
Turbidity value with dark offset and
gain applied.
TURB COUNTS
40038
Unsigned Integer
2
R
Turbidity A/D counts
TEMP MAX
40040
Float
2
R
Maximum temperature
TEMP MIN
40042
Float
2
R
Minimum temperature
LAMP V
40044
Float
2
R
Lamp voltage
LAMP CURR
40046
Float
2
R
Lamp current (amps)
Plus 5V
40048
Float
2
R
Plus five volt measurement
INPUT V
40050
Float
2
R
Input voltage (~12V)
VREF
40052
Float
2
R
Voltage reference measurement (2.5V)
CAL GAIN
40067
Float
2
R
Calibration gain factor - used to convert A/D
counts to turbidity
INITIALS
40083
String
2
R
Initials used for the latest calibration
LAST CAL DATE
40085
Time2
2
R
Time of the latest calibration
CAL VALUE
40087
Float
2
R
The standard value used for the latest
calibration