Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual
Page 90
85
3.1
Offsets – Input offsets
3.2
CJ gain – Gain of cold-junction compensation
3.3
TC neg gain – Thermocouple negative-gain slope
3.4
TC pos gain – Thermocouple positive-gain slope
3.5
CJ Correction – Offset of cold-junction compensation
3.6
DAC coarse gain – Coarse output DAC gain slope
3.7
DAC fine gain – Fine output DAC gain slope
3.8
DAC offset – DAC offset constant
3.9
Full Recal – Sequences from step 3.1 to step 3.8 and stores all calibration constants.
4.0 Specifications
Input impedance: > 10 megohms
Warm-up time:
≤ 1 Minute to full rated accuracy
mV Accuracy:
± 0.025% R + 0.005% FS
Temperature Effect:
≤ 2uV/°C
T/C Accuracy:
± 0.3°C for E, J, K, N & T types
T/C Accuracy:
± 1.0°C for B, R & S types
T/C Accuracy:
±0.5°C for C, D, M & P types
Cold junction effect:
± 0.5°C max
Resolution:
± 0.1° or ± 1° for T/Cs, ± 0.001 mV
Voltage range:
± 110.000 mV dc
Temperature: Storage: -40
° F to 140° F (-40° C to 60° C)
Operating: 23
° F to 122° F (-5° C to 50° C)
Connector – Measure & Simulate T: standard mini-t/c connector with t/c wire
- Source mV: standard mini-t/c connector with cable and two alligator clips
Power: supplied from MFT (simulate / source mode only)
Units:
°F, °C, °R, °K, milli-Volts
Weight: 3 ozs
5.0 Approvals
5.1 Intrinsically Safe by MET Laboratories, Inc per CAN/CSA-C22.2 No. 157, CAN/CSA-C22.2 No. 1010.1-92
& UL 913 Fifth Edition, Rev 2/21/97 for Class I, Division I, Groups A, B, C &D Hazardous Areas.
Restrictions apply – see the Intrinsically Safe Control Document in Appendix for more details.
5.2 CE mark
6.0 Accessories for TIO0110
P/N A36857
T/C Wire Kit 1 for types E, J, K, N, T - mini t/c connector, 1 m cable, wire ends
P/N A36858
T/C Wire Kit 2 for types B, K, R - mini t/c connector, 1 m cable, wire ends
P/N A36881
mV simulation cable assembly; mini t/c connector, 3.5 ft. copper cable, alligator clips