Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 95

Advertising
background image

90

Start – Begins the desired simulation mode with the stepping

parameters selected.

Mode – Displays the current selected simulation mode. Select

Mode to change.

Low pt – Displays the starting source or simulation value.

Select to enter a value.

Upr pt – Displays the ending source or simulation value. Select

to enter a value.

Steps – Displays the number of steps to take (including the

“Upr pt”). Select to change the number of steps.

Dwell Time – Displays the time delay in seconds at each step.

“Manual” can be selected to enable user to
manually advance to next programmed step.




5.0 Resolution, Range, Accuracy for VMA0055 Module

mA dc

mA Measure Mode

mA Source Mode

Resolution

.001 mA

.001 mA

Range

±100.000 mA (55 Vdc compliance)

0 - 22.000 mA

Accuracy

± (0.01% of reading + 0.015% FS)

± (0.01% of reading + 0.015% FS)

Open Circuit Voltage

24 VDC

Output Drive

15 VDC minimum @ 24 mA, Resistive load

Volts dc

V dc Measure Mode

V dc Source (Regulated) Mode

Resolution

1/100,000 counts: .001 mV; .001 V

1 / 1,000,000

Ranges

500 mV; 1, 2, 4, 8, 15, 30, 55 V

± 1000 mV, 0 - 24.000 VDC

Accuracy

± (0.01% of reading + 0.05% FS)

± (0.01% of reading + 0.05% FS)

Open Circuit Voltage

24 VDC

Output Drive

15 VDC minimum @ 24 mA, Resistive load

2-wireTransmitter Simulate

mA Simulation

Resolution

.001 mA

Range

0 - 24.000 mA

Accuracy

± (0.01% of reading + 0.015% FS)

Loop Voltage Limits

1 VDC min., 55 VDC max.

Regulated Loop Power

Regulated Power

Resolution

na

Range

24V dc

Accuracy

± (0.01% of reading + 0.015% FS)

Open Circuit Voltage

24 VDC

Output Drive

15 VDC minimum @ 24 mA, Resistive load


Notes: 1. Nominal resistance at VMA current terminal is 10 – 15 Ω
2. Output load line is linear

Auto-Step Setup Display

Vma Setup
Auto-Step 1:
Start
Mode:

mA Source

Low pt:

4.000 mA

Upr pt:

20.000 mA

Steps:

5

Auto-Step 3
Dwell Time: Manual


Up Down Select Back

Advertising