Multichannel Systems TC01-2 Manual User Manual

Page 16

Advertising
background image

Temperature Controller TC01/02 Manual

12

3.3.6 Hardware

Diagnosis

This menu should be used for reviewing parameter settings or check the hardware performance
in case that you observe any trouble with the instrument. Each channel can be checked separately.
If the problem persists, please contact your local retailer. The highly qualified staff will be glad to
help you. Keep the displayed information at hand when contacting customer support.

There are four separate screen views with different sets of information in the Diagnosis menu.
You can toggle between the view by pressing the Up and Down buttons.

Diagnosis 1: Measured values

This diagnose screen view is used for checking the temperature sensor.

Temperature

Actual temperature

Resistance 2

Cable resistance high side of sensor, see also chapter "D-Sub9 Pin
Assignment".

Resistance 1

Cable resistance low side of sensor, see also chapter "D-Sub9 Pin
Assignment".

Resistance X

Sensor resistance plus cable resistance

Resistance S

Sensor resistance

Board Temp

Board temperature (The TC01/02 will switch off the channel
outputs and go into stand-by mode when the board temperature
reaches 90 °C.)

Diagnosis 2: Controller settings

This diagnose screen view is used for reviewing and checking the user settings.

Setpoint Temp

Setpoint temperature

P Gain

Proportional gain

I Gain

Integrator gain

Max Power

Maximum output power

Diagnosis 3: Controller output

This diagnose screen view is used for checking the operation of the internal controller.

Power Set

Output power set by the controller.

Power Out

Actual output power (product of Current Out and Supply
Voltage
)

Duty Cycle

PWM Duty cycle (internal value)

Current Out

Current through the primary coil of the isolating transformer

Supply Voltage

Supply voltage (from power supply)