Installation and Operation

13


Diagnosis 4: Heating element

This diagnose screen view is used for checking the connected heating element.

On/off

Current channel state

HE Voltage

Output voltage applied to heating element

HE Current

Output current applied to heating element

HE Resistance

Heating element resistance (voltage-current ratio)

HE Power

Output power delivered to heating element (voltage-current
product), should be 80-90 % of Power Out, depending on heating
element resistance)