Fairbanks X SERIES PR 5410 Process Indicator User Manual

Page 94

Advertising
background image

Section 4: Commissioning

10/08 94

51207 Revision 1

4.8.2 Configuring Digital Inputs and Outputs

Press

and select [Digital I/O parameter] to open the configuration menu for outputs and inputs:

Setup/Digital I/O parameter

Output 1

marker bit 1

X64=1

Output 2

marker bit 2

X65=1

Output 3

marker bit 3

X66=1

Input 1 on

-no action-

Input 1 off

-no action-

Input 2 on

-no action-

Input 2 off

-no action-

Input 3 on

-no action-

Input 3 off

-no action-

BCD out

Gross


4.8.3 Configuring Outputs


Configure the required function for [Output 1] to [Output 3] by selecting a signal from the list; see
Chapter 4.8.1. The output goes to the corresponding state.

Example:

Setup/Digital I/O parameter

Output 1

below zero

X35=1

Output 2

above Max

X33=0

Output 3

center zero

X36=1

Input 1 on

-no action-


[Output 1] is true (active), when the weight value drops below zero (X35=1).
[Output 2] remains (active), as long as the weight is not above Max (X33=0).
[Output 3] is true (active), when the weight is zero +/- 1/4 d (X36=1).