4 oem molex connector (e0 version), Fig. 1-2 bnc connector, Fig. 1-3 molex connector pin-out in e0 version – Gentec-EO UP and XLP12 User Manual

Page 7

Advertising
background image

Version 5.0

Ultra Series UP Instruction Manual

Gentec Electro-Optics Inc. All rights reserved

3

1.2.2 Integra USB connector

The Integra USB Connector is an integrated monitor that allows to plug the head directly into a
computer. It has the same serial commands as the MAESTRO and a few extra

one’s (see the PC-

Gentec-EO Manual) and uses the same PC-Gentec-EO software. All specifications are mostly the
same, except for:

Wattmeter and joulemeter heads have a noise ~1.3 x higher.
The risetime of the UP12 head (with anticipation) is slower by 0.2 seconds.




1.2.3 BNC connector


The BNC connector is quick and easy to install and is the best at shielding EMI noise.

The BNC connector allows you to connect the detector directly on an oscilloscope or to a precision
microvoltmeter with the correct load impedance. The connector is available in the UP detectors
without amplification (load impedance: 100 k

Ω) and with the Molex connector in the amplified

version (load impedance : 10 M

Ω).


FIG. 1-2 BNC CONNECTOR



1.2.4 OEM Molex connector (E0 version)

The Molex female connector is available in the UP19-xxx-H5 and UP12E-xxx-H5 without
amplification (load impedance : 100 k

Ω). It allows you to connect to an external PCB board

available at Gentec-EO.

FIG. 1-3 MOLEX CONNECTOR PIN-OUT IN E0 VERSION

Advertising