Error between channels test, Excessive input test – Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 13

Advertising
background image

2-4

SM 04L31A01-01E

Measurement Accuracy Test (Measurement Input and Control Input)

Check that the specifications below are met given the following operating conditions: 23

±

2

°

C, 55

±

10% RH, and a warm up time of 30 minutes.

Range

Input Value

Criterion

Specifications

20 mV

+20.00 mV

±

3 digits

±

(0.1% of rdg + 2 digits)

0.00 mV

±

2 digits

–20.00 mV

±

3 digits

60 mV

+60.00 mV

±

6 digits

±

(0.1% of rdg + 2 digits)

0.00 mV

±

2 digits

–60.00 mV

±

6 digits

200 mV

+200.0 mV

±

3 digits

±

(0.1% of rdg + 2 digits)

0.0 mV

±

1 digits

–200.0 mV

±

3 digits

2 V

+2.000 V

±

3 digits

±

(0.1% of rdg + 2 digits)

0.000 V

±

1 digits

–2.000 V

±

3 digits

6 V

+6.000 V

±

6 digits

±

(0.1% of rdg + 2 digits)

0.000 V

±

1 digits

–6.000 V

±

6 digits

20 V

+20.00 V

±

3 digits

±

(0.1% of rdg + 2 digits)

0.00 V

±

1 digits

–20.00 V

±

3 digits

50 V

+30.00 V

±

5 digits

±

(0.1% of rdg + 3 digits)

0.00 V

±

2 digits

–30.00 V

±

5 digits

Pt100

18.52

/–200

°

C

±

0.4

°

C

±

(0.15% of rdg + 0.3

°

C)

100.00

/0

°

C

±

0.2

°

C

313.71

/600

°

C

±

0.9

°

C

Error between Channels Test

Connect the (+), (–), and (b) terminals on all measurement and control input channels

using thick leads, then check whether the following criteria are met.

Range

Input Value

Criterion

20 mV

Short (+) (–)

0.00 mV

±

2 digits

Pt100

103.9

Measured value: 10

°

C

±

0.2

°

C,difference between channels:

0.2

°

C or less

Note

If the measurement error does not meet the criterion (10

°

C

±

0.2

°

C), and unless you

determine the cause to be the effect of the resistance in the test facility's RTD wiring, use

the following measurement criteria.

Measured value:

10

°

C

±

0.5

°

C

Difference between channels: 0.2

°

C or less

See your nearest Yokogawa representative for the error between channels in the control

input section.

Excessive Input Test

1.

Connect the plus (+) and minus (–) sides of all measurement and control input

channels.

2.

Set all channels to the 20 mV range.

3.

Apply 10 V between the plus and minus sides for 1 minute, and confirm that

there is no malfunction.

4.

Set all channels to the Pt100 range.

5.

Apply

±

10 V between (A) and (B) for 1 minute, and confirm that there is no

malfunction.

2.2 Test Procedures

Advertising