Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 27

Advertising
background image

3-7

SM 04L31A01-01E

Adjustments

3

6.

Press the Write function key for each output number. The calibration values are

saved.

7.

Repeat steps 4—6 and calibrate all module and output numbers.

8.

Press ESC. You are returned to the calibration screen.

9.

Turn OFF the power

Calibration Using Communications

The commands required for control output calibration are listed below.

For other necessary commands, see Calibration Using Communications (page 3-5) in

section 3.1, “Calibrating the Measuring Instrument Input.”

Perform step 1 of Calibration Using Keys before using these commands for

calibration.

For OUTPUT

ZZp1,p2,p3,p4

p1

Action OUTPUT

p2

Module number (1 for the CX1000, or 1—3 for the CX2000)

p3

Output channels (1 and 2)

p4

4 mA/20 mA

p5

Calibration value (For 4 mA, 9285—12561)

(For 20 mA, 46421—62804)

For WRITE

ZZp1,p2,p3,p4

p1

Action (OUTPUT/WRITE)

p2

Module number (1 for the CX1000, or 1—3 for the CX2000)

p3

Output channels (1 and 2)

p4

(Calibration value for 4 mA, 9285—12561)

p5

(Calibration value for 20 mA, 46421—12561)

Note

When specifying WRITE, you can omit p4 and p5. If you do so, the currently set calibration

value is written.

3.2 Control Output Calibration