4.3
Specifications of mA-input Module
Style Number: S5
Model Code, Number of Input Channels, Terminal Configuration and Shortest
Measurement Interval
Model Code
Number of Channels Terminal Configuration
Shortest Measurement Interval
DU300-11
10
Screw
0.5 s
DU300-12
10
Clamp
0.5 s
Method of Input
Non-balanced floating input with isolation between channels (separated channels)
Resolution of A/D Conversion
±20000
Integral Time of A/D Conversion
Manual or automatic selection between 20 ms (50 Hz), 16.7 ms (60 Hz) and 100 ms (10 Hz)
However, AUTO does not function if the instrument is the stand-alone DC power supply model,
or the subunit (DS400/DS600) of the DC power supply model. (Selecting “AUTO” will set the
A/D integration time to 20 ms (50 Hz)).
Shortest Measurement Interval and Cutoff Frequency
When the lowpass filter is turned off:
Cutoff frequency:
50/60 Hz
10 Hz
Shortest measurement interval:
0.5 s*
4 s
*: 2s if the power monitor module is installed
When the lowpass filter is turned on:
Cutoff frequency:
50/60 Hz
10 Hz
Shortest measurement interval:
3 s
12 s
Withstanding Voltage
Across channels: 100 VA C, 1 min
Across output terminals and ground: 1,500 V AC, 1 min (excluding DU500-14)
Normal Operating Temperature/Humidity Range
-10˚ to 60˚C
20 to 80% RH for an ambient temperature range of -10˚ to 50˚C; 10 to 50% RH for 40˚ to 50˚C; 5
to 30% RH for 50˚ to 60˚C (non-condensing)
Maximum Input Voltage
5 V DC
Maximum Normal-mode Noise Current
24 mA (peak value, including 50- and 60-Hz signal components)
Equivalent voltage of 2.4 V
Normal Mode Rejection Ratio (Equivalent Voltage)
40 dB minimum (50/60 Hz
±0.1%)
Maximum Common Mode Noise Voltage
250 V ACrms (50/60 Hz)
Common Mode Rejection Ratio (Equivalent Voltage)
120 dB minimum (50/60 Hz
±0.1%)
Maximum Noise Across Channels
150 V ACrms (50/60 Hz)
Measuring Range/Accuracy
Measured under the following standards of operating conditions:
Reference operation conditions
temperature: 23
±2°C; humidity: 55 ±10% RH; supply voltage 90 to 250 VAC (AC power supply)
/10 to 32 VDC (DC power supply); supply frequency: 50/60
±1% (AC power supply); warm-up
time: at least 30 minutes; when operating, the system must not adversely affect the operation of
other equipment.
Measuring Range
Measuring Accuracy
Maximum
Resolution
-20.000 to 20.000 mA
±(0.25% of rdg. + 5 digits)
1
µA