3 specifications of ma-input module, Specifications of ma-input module -12 – Yokogawa DA100 User Manual

Page 81

Advertising
background image

IM DA100-01E

4-12

4.3

Specifications of mA-input Module

Style Number: S5
Model Code, Number of Input Channels, Terminal Configuration and Shortest
Measurement Interval

Model Code

Number of Channels Terminal Configuration

Shortest Measurement Interval

DU300-11

10

Screw

0.5 s

DU300-12

10

Clamp

0.5 s

Method of Input

Non-balanced floating input with isolation between channels (separated channels)

Resolution of A/D Conversion

±20000

Integral Time of A/D Conversion

Manual or automatic selection between 20 ms (50 Hz), 16.7 ms (60 Hz) and 100 ms (10 Hz)
However, AUTO does not function if the instrument is the stand-alone DC power supply model,
or the subunit (DS400/DS600) of the DC power supply model. (Selecting “AUTO” will set the
A/D integration time to 20 ms (50 Hz)).

Shortest Measurement Interval and Cutoff Frequency

When the lowpass filter is turned off:

Cutoff frequency:

50/60 Hz

10 Hz

Shortest measurement interval:

0.5 s*

4 s

*: 2s if the power monitor module is installed

When the lowpass filter is turned on:

Cutoff frequency:

50/60 Hz

10 Hz

Shortest measurement interval:

3 s

12 s

Withstanding Voltage

Across channels: 100 VA C, 1 min
Across output terminals and ground: 1,500 V AC, 1 min (excluding DU500-14)

Normal Operating Temperature/Humidity Range

-10˚ to 60˚C
20 to 80% RH for an ambient temperature range of -10˚ to 50˚C; 10 to 50% RH for 40˚ to 50˚C; 5
to 30% RH for 50˚ to 60˚C (non-condensing)

Maximum Input Voltage

5 V DC

Maximum Normal-mode Noise Current

24 mA (peak value, including 50- and 60-Hz signal components)
Equivalent voltage of 2.4 V

Normal Mode Rejection Ratio (Equivalent Voltage)

40 dB minimum (50/60 Hz

±0.1%)

Maximum Common Mode Noise Voltage

250 V ACrms (50/60 Hz)

Common Mode Rejection Ratio (Equivalent Voltage)

120 dB minimum (50/60 Hz

±0.1%)

Maximum Noise Across Channels

150 V ACrms (50/60 Hz)

Measuring Range/Accuracy

Measured under the following standards of operating conditions:

Reference operation conditions

temperature: 23

±2°C; humidity: 55 ±10% RH; supply voltage 90 to 250 VAC (AC power supply)

/10 to 32 VDC (DC power supply); supply frequency: 50/60

±1% (AC power supply); warm-up

time: at least 30 minutes; when operating, the system must not adversely affect the operation of
other equipment.

Measuring Range

Measuring Accuracy

Maximum

Resolution

-20.000 to 20.000 mA

±(0.25% of rdg. + 5 digits)

1

µA