SERVICE

7

57

UFM 530 HT

www.krohne.com

12/2013 - 4002393201- HB UFM 530 HT R02 en

7.2 Maintenance

Contact your local representative for maintenance or repair.

7.2.1 Admonitions for replacement of flow sensor in separate systems

Specific calibration data for each flow sensor are determined during factory calibration. Flow
sensor specific data must be set in the UFC 030 ultrasonic flow converter when a flow sensor is
replaced.

• Enter value of flow sensor primary constant GK in menu 3.01.09 (it is specified on the

nameplate).

• If the new flow sensor has a different meter size, you have to program this menu 3.1.8.
• Also set the full-scale range for Q

100%

in menu 3.1.1.

• Make a zero check after setting of new data.

7.3 Cleaning

DANGER!
The following instructions are for use by qualified personnel only. To avoid electric shock, do not
perform any servicing other than described in the operating instructions of this manual unless
you are qualified to do so.

DANGER!
Always switch off power source before starting work!

CAUTION!
Scaling at the innerside of the pipe effects the flow measurement. Clean the pipe in such case.

INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin-
free and acid-free grease.
Ensure that the housing gasket is properly fitted, clean and undamaged.