Fct. 1.1.4 TIMECONST. time constant for flow rate

Enter a low-pass filter time constant in seconds to be
applied to flow rate. A value of zero indicates that
low-pass filter is not to be applied. With this function
it is possible to compromise between a steady
indication (on display/current output) and response
time (to flow changes).

LIMITS - 0 to 20 seconds.

APPEARS - Always.

Fct. 1.2.0 DISPLAY

This submenu groups the display functions which

- Allow selection of units for all measured parameters.
- Select what parameters to include in display cycle.
- Select display mode (cyclic/non-cyclic) and error
messages to appear or not to appear in display cycle.

APPEARS - Alwayss

Fct. 1.2.1 FLOW UNITS for display

Select a unit in which flow rate is to be displayed

from the following list of the available units,

depending on programming of Fct.1.1.1MEAS.

INST.

for volumetric flow -
l m3/hr

l m3/min

l m3/sec

l Litre/hr

l Litre/min

l Litre/sec

l ft3/hr

l ft 3/min

l ft3/sec

l cft/hr

l cft/min

l cft/sec

l cuft/hr

l cuft/min

l cuft/sec

l US Gal/hr

l US Gal/min l US Gal/sec

l UK Gal/hr

l UK Gal/min l UK Gal/sec

l % MAX. FLOW
for normalized volumetric flow -

l Norm.m3/hr l Norm.m3/ min l Norm.m3/ sec

l Norm.L/hr

l Norm.L/min

l Norm.L/sec

l Sft3/hr

l Sft3/min

l Sft3/sec

l Scft/hr

l Scft/min

l Scft/sec

l % MAX. FLOW
For mass flow -

l Kg/hr

l kg/min

l kg/sec

l T/hr

l T/min

l T/sec

l Lb/hr

l Lb/min

l Lb/sec

l % MAX. FLOW
Note that the list is same as for Fct. 1.1.2 except for
an additional unit % MAX. FLOW (to display flow
rate as a percentage of max. flow).

LIMITS - Not applicable.
APPEARS - Always.

Fct. 1.2.2 TOTAL.UNITS totalizer unit for display

Totalized flow may be displayed in one of the
following units.

For volumetric flow -

l m3

l Litre

l ft3

l cft

l cuft

l US Gal

l UK Gal

l NO DISPLAY

For normalized volumetric flow -

l Norm.m3

l Norm.L

l Sft3

l Scft

l NO DISPLAY

For mass flow -

l kg

l T

l Lb

l NO DISPLAY
Use NO DISPLAY to exclude totaliser for display
cycle.

LIMITS - Not applicable.

APPEARS - Always.

Fct. 1.2.3 TEMP. UNITS for display

The following options exist for temperature unit

l Deg. C

l Deg. F

l KELVIN

l NO DISPLAY

Use NO DISPLAY to exclude temperature for
display cycle.

LIMITS - Not applicable.

APPEARS - Always.

Fct. 1.2.4 PRES. UNITS pressure unit for display

The following options exist for pressure. Unit with the
suffix _g are gauge pressure units and those without
the same are absolute pressure units.

l Pa

l KPa

l atm

l Bar

l mBar

l PSI

l Lbf/ft2

l Kg/cm2

l In. Hg

l mm Hg

l mm Water

l pa_g

l KPa_g

l atm_g

l Bar_g

l mBar_g

l PSI_g

l Lbf/ft2_g

l Kg/cm2_g

l In. Hg_g

l mm Hg_g

l mm Water_g l NO DISPLAY
Use NO DISPLAY to exclude pressure from display
cycle.

LIMITS - Not applicable.

APPEARS - If Fct. 3.5.1 FLUID is not LIQUID.