6

3.3 Outputs

3.3.1 Abbreviations

Abbreviation Stands

for Programming

via Fct. No...

Description

See Sect....

EC
EMC
F
F

100%

I
I

0%

I

100%

Q

0%

Q

100%

Electronic counter
Electro-mechanical counter
Frequency (Pulse) output
Pulses for Q = 100 % flow rate or pulse value
Current (analog) output
Current at 0/4 mA flow
Current at 20 mA flow
0% flow rate
Full - Scale range, 100 % flow rate.

---
---
1.4.X
1.4.2
1.3.X
1.3.3
1.3.4

1.1.2

---
---
6.1 + 6.2.9
6.1 + 6.2.9
6.1 + 6.2.8
6.1 + 6.2.8
6.1 + 6.2.8
-
6.1 + 6.2.4


3.3.2 Current (analog) output I

The current output is galvanically isolated from all input and output circuits but not from frequency

output F. Therefore only one grounded receiver instrument may be connected to either current output I or
frequency output F.

·

·

·

·

All functions and operating data are programmable, see sections 5, 6.1 & 6.2.8.
Factory-set data and functions are listed in the enclosed ‘Configuration Sheet’ on settings.

This can also be used to record any changes made to the operating parameters.

Max load at terminals 5/6 for I

100%

(Fct. 1.3.4) :

Max. load Kohms =

]

[

14

%

100

mA

V

(e.g. 0.7 K ohms for I

100%

= 20 mA)

[

]

I

Error annunciation programmable to 2 mA or 22 mA (Fct. 1.3.2)

·

·

Connection diagram

 Refer to Section 3.3.4