Technical data – KROHNE OPTIWAVE 6300C EN User Manual

Page 84

Advertising
background image

8

TECHNICAL DATA

84

OPTIWAVE 6300 C

www.krohne.com

03/2014 - 4000547004 - HB OPTIWAVE 6300 R04 en

Process connections

Thread

G 1½ (ISO 228); 1½ NPT (ASME B1.20.1)

Flange version

Flange version

Flange version

Flange version

EN 1092-1

DN80 in PN40 (Form B1), DN100…200 in PN16 or PN40 (Form B1); others on

request

ASME B16.5

3¨…8¨ in 150 lb RF, 3¨...4¨ in 300 lb RF; others on request

JIS B2220

80…100A in 10K; others on request

Other

Others on request

Electrical connections

Power supply

Terminals output 1

Terminals output 1

Terminals output 1

Terminals output 1 Non-Ex / Ex i:

Non-Ex / Ex i:

Non-Ex / Ex i:

Non-Ex / Ex i:

14…30 VDC; min./max. value for an output of 22 mA at the terminal

Terminals output 1

Terminals output 1

Terminals output 1

Terminals output 1 Ex d:

Ex d:

Ex d:

Ex d:

20…36 VDC; min./max. value for an output of 22 mA at the terminal

Terminals output 2

Terminals output 2

Terminals output 2

Terminals output 2 Non-Ex / Ex i / Ex d:

Non-Ex / Ex i / Ex d:

Non-Ex / Ex i / Ex d:

Non-Ex / Ex i / Ex d:

10…30 VDC; min./max. value for an output of 22 mA at the terminal (additional

power supply needed – output only)

Cable entry

M20×1.5; ½ NPT

G ½ (not for FM- and CSA-approved devices. Not for stainless steel housings.)

M25×1.5 (for stainless steel housings only)

Cable gland

Standard: none

Options: M20×1.5 (for non-Ex and Ex -approved devices with M20×1.5 and M25×1.5

cable entries); others are available on request

Cable entry capacity (terminal)

0.5…1.5 mm²

Input and output

Current output

Current output

Current output

Current output

Output signal

(Output 1)

4…20 mA HART

®

or 3.8…20.5 mA acc. to NAMUR NE 43

4

Output signal

(Output 2 - optional)

4…20 mA (no HART

®

signal) or 3.8…20.5 mA acc. to NAMUR NE 43

Resolution

±3 µA

Temperature drift

Typically 25 ppm/K

Error signal

High: 22 mA; Low: 3.6 mA acc. to NAMUR NE 43

Advertising