Appendix, Appendix 7.1 orp as a function of mv – KROHNE OPTISENS ORP 8590 EN User Manual

Page 30

Advertising
background image

7

APPENDIX

30

OPTISENS ORP 8590

www.krohne.com

01/2014 - 4001933801 MA OPTISENS ORP 8590 R01 en

Appendix

7.1 ORP as a function of mV

The pH-value is the negative decadative logarithm of the hydrogen ion concentration, and it is
directly related to the proportion of hydrogen ions H

+

to hydroxide ions OH

-

in the media. The pH-

sensor measures excess or deficit of the hydrogen ions and gives a proportional millivolt signal
as output. The signal is 59.16 mV per 1 pH at 25°C / 77°F. In clean water there is a total balance
between hydrogen ions and hydroxide ions, the output from the electrode is 0.0 mV and pH is 7.
The millivolt signal is measured by the pH sensor and the corresponding pH value is calculated
in the signal converter.

mV

pH

H

+

ions [mol/l]

OH

-

ions [mol/l]

Example

414

0

1

0.00000000000001

355

1

0.1

0.0000000000001

296

2

0.01

0.000000000001 Coca Cola

237

3

0.001

0.00000000001

177

4

0.0001

0.0000000001 Orange juice

118

5

0.00001

0.000000001

59

6

0.000001

0.00000001 Milk

0

7

0.0000001

0.0000001 Clean water

-59

8

0.00000001

0.000001 Blood

-118

9

0.000000001

0.00001

-177

10

0.0000000001

0.0001

-237

11

0.00000000001

0.001

-296

12

0.000000000001

0.01

-355

13

0.0000000000001

0.1

-414

14

0.00000000000001

1 Sulfa

.book Page 30 Friday, January 24, 2014 11:29 AM