Appendix, 2 orp dependency – KROHNE OPTISENS ORP 8590 EN User Manual

Page 31

Advertising
background image

APPENDIX

7

31

OPTISENS ORP 8590

www.krohne.com

01/2014 - 4001933801 MA OPTISENS ORP 8590 R01 en

7.2 ORP dependency

The output from a pH-electrode varies with the temperature in a predictable way. The size of the
variation depends on both the temperature and the pH being measured.

At pH 7 and 25°C / 77°F the temperature error is zero. If temperature or pH changes the
temperature error is calculated using the following formula:

0.03 pH-difference / pH or 0.03 pH-difference / K

°

C

°

F

pH

pH

pH

pH

pH

pH

pH

pH

pH

pH

pH

pH

5

41

2.30

3.24

4.18

5.12

6.06

7.00

8.06

9.12

10.18 11.24 12.30 13.36

15

59

2.15

3.12

4.09

5.06

6.03

7.00

8.03

9.06

10.09 11.12 12.15 13.18

25

77

2.00

3.00

4.00

5.00

6.00

7.00

8.00

9.00

10.00 11.00 12.00 13.00

35

95

1.85

2.88

3.91

4.94

5.97

7.00

7.97

8.94

9.91

10.88 11.85 12.82

45

113

1.70

2.76

3.82

4.88

5.94

7.00

7.94

8.88

9.82

10.76 11.70 12.64

55

131

1.55

2.64

3.73

4.82

5.91

7.00

7.91

88.2

9.73

10.64 11.55 12.46

65

149

1.40

2.52

3.64

4.76

5.88

7.00

7.88

8.76

9.64

10.52 11.40 12.28

75

167

1.25

2.40

3.55

4.70

5.85

7.00

7.85

8.70

9.55

10.40 11.25 12.10

85

185

1.10

2.28

3.46

4.64

5.82

7.00

7.82

8.64

9.46

10.28 11.10 11.92

95

203

0.95

2.16

3.37

4.58

5.79

7.00

7.79

8.58

9.37

10.16 10.95 11.74

.book Page 31 Friday, January 24, 2014 11:29 AM

Advertising