3 offset calibration – KROHNE OPTISENS PAM 2080 EN User Manual

Page 36

Advertising
background image

www.krohne.com

36

6.3 Offset Calibration

Offset calibration is used to compensate for the electrode drift that sets in due to the interaction of the

probes with the measuring medium.

This drift is dependent upon the substances contained in the water and is therefore different for every

measuring point. With time, you will establish how often you need to carry out an offset in each case.

Offset calibration is very easy to carry out. The probe generally remains in the measuring medium in the

process.

Carrying Out Offset Calibration

Select the measuring channel that you wish to calibrate.

Set the calibration type to offset calibration ("Calibration – Cal.method – Offset). Ensure that

measurement is active (must not remain on “Hold”).

Now press the Cal button once.


Displays of “Hold”, “Electrode active”, “Buffer”, “Run time” indicate that the offset routine is

running.

The symbols “Hold”, “Buffer” and “Run time” are displayed after approx. 20 sec.

Now press the Cal button once more in order to save the determined measurement internally.

You now have 2 hours.

Advertising
This manual is related to the following products: