Operation – KROHNE OPTIWAVE 5200 FOUNDATION FIELDBUS User Manual

Page 30

Advertising
background image

4

OPERATION

30

OPTIWAVE 5200 C

www.krohne.com

09/2013 - 4002500101 - AD FF OPTIWAVE 5200 R01 en

FD_OFFSPEC_MASK

FD_OFFSPEC_MASK

FD_OFFSPEC_MASK

FD_OFFSPEC_MASK

Offspec Mask

Offspec Mask

Offspec Mask

Offspec Mask

RW

This parameter allows the user to suppress any single or

multiple conditions that are active, in this category, from

being broadcast to the host through the alarm parameter.

Nothing masked

Setting:

(00) Check Function

(01) Cond01

(02) Cond02

(03) Cond03

(04) Cond04

(05) Cond05

(06) Cond06 disabled

(07) Bad Measurement quality disabled

(08) Signal Strong disabled

(09) Signal Weak disabled

(10) Empty Spectrum invalid disabled

(11) Sensor No Signal

(12) Sensor Microwave Error disabled

(13) Temperature Above +75°C disabled

(14) Temperature Below -35°C disabled

(15) Overfill Warning disabled

(16) Level Lost warning disabled

(17) Temperature out of range (warning) disabled

(18) Local operation disabled

(19) Overfill Error disabled

(20) Level lost error disabled

(21) Temperature out of range Error disabled

(22) Converter Voltage Error disabled

(23) Converter ROM Error disabled

(24) Converter RAM Error disabled

(25) Converter EEPROM Error disabled

(26) Sensor not compatible disabled

(27) Internal Communication Error disabled

(28) Sensor Voltage Error disabled

(29) Sensor ROM Error disabled

(30) Sensor RAM Error disabled

(31) Sensor EEPROM Error disabled

FD_OFFSPEC_PRI

FD_OFFSPEC_PRI

FD_OFFSPEC_PRI

FD_OFFSPEC_PRI

Offspec Priority

Offspec Priority

Offspec Priority

Offspec Priority

RW

This parameter allows the user to specify the priority of this

alarm category.

0

Setting:

0...15

Parameter

Parameter

Parameter

Parameter
DD name

Access

Description and settings

Initial Value

Advertising