3 scaling the analogue outputs – Vaisala HMP260 User Manual
Page 39
Advertising
HMP260 SERIES
HMP260-U017en-1.5
Operating Manual
1997-11-25
33
5.4.1.2
Selecting and scaling the analogue output quantities
Disconnect the security lock jumper!
ASEL xxx yyy <cr>
xxx
=
channel 1's quantity
yyy
=
channel 2's quantity (RH, T, Td, Abs, Mix or Tw)
Example:
relative humidity selected on channel 1 and temperature on channel 2
ASEL RH T <cr>
Ch1 (RH)
lo
0.000 %RH ? <cr>
Ch1 (RH)
hi
100.000 %RH ? <cr>
Ch2 (T )
lo
-40.000 'C
? <cr>
Ch2 (T )
hi +160.000 'C
? <cr>
5.4.1.3
Scaling the analogue outputs
Disconnect the security lock jumper!
ASCL <cr>
Example:
relative humidity is scaled on the range of 0...100 %RH and temperature -
40...+160 °C
ASCL <cr>
Ch1 (RH)
lo
0.000 %RH ? <cr>
Ch1 (RH)
hi 100.000 %RH ? <cr>
Ch2 (T )
lo
0.000 'C
? -40 <cr>
Ch2 (T )
hi 100.000 'C
? 160 <cr>