6 selecting the output units, 7 setting the transmitter address, 8 resetting the transmitter – Vaisala HMP260 User Manual
Page 42: 3 operating the transmitter via the serial bus, 1 setting the serial interface
Advertising
HMP260 SERIES
Operating Manual
HMP260-U017en-1.5
36
1997-11-25
5.4.2.6
Selecting the output units
UNIT x <cr>
x
=
m(etric units)
n(on-metric units)
metric
non-metric
RH
%RH
%RH
T
°C
°F
Td
°C
°F
a
g/m
3
gr/ft
3
x
g/kg
gr/lb
Tw
°C
°F
5.4.2.7
Setting the transmitter address
ADDR aa <cr>
aa
=
address (0...99)
Example:
transmitter is given address 99
ADDR <cr>
Address : 2 ? 99 <cr>
5.4.2.8
Resetting the transmitter
RESET <cr>
5.4.3
Operating the transmitter via the serial bus
5.4.3.1
Setting the serial interface
SMODE xxxx<cr>
xxxx
=
STOP, RUN or POLL
In STOP mode: measurements output only by command, all commands can be
used
In RUN mode:
outputting automatically, only command S can be used