6 selecting the output units, 7 setting the transmitter address, 8 resetting the transmitter – Vaisala HMP260 User Manual

Page 42: 3 operating the transmitter via the serial bus, 1 setting the serial interface

Advertising
background image

HMP260 SERIES
Operating Manual

HMP260-U017en-1.5

36

1997-11-25

5.4.2.6

Selecting the output units

UNIT x <cr>

x

=

m(etric units)
n(on-metric units)

metric

non-metric

RH

%RH

%RH

T

°C

°F

Td

°C

°F

a

g/m

3

gr/ft

3

x

g/kg

gr/lb

Tw

°C

°F

5.4.2.7

Setting the transmitter address

ADDR aa <cr>

aa

=

address (0...99)

Example:
transmitter is given address 99

ADDR <cr>
Address : 2 ? 99 <cr>

5.4.2.8

Resetting the transmitter

RESET <cr>

5.4.3

Operating the transmitter via the serial bus

5.4.3.1

Setting the serial interface

SMODE xxxx<cr>

xxxx

=

STOP, RUN or POLL

In STOP mode: measurements output only by command, all commands can be

used

In RUN mode:

outputting automatically, only command S can be used