List of ptb330 serial commands, Figure 27, Female m12 connector with screw terminals – Vaisala PTB330TS User Manual

Page 68

Advertising
background image

User's Guide ______________________________________________________________________

66 ___________________________________________________________________M210983EN-A

A female M12 connector with screw terminals is provided with the
PTB330TS; see

Figure 27 on page 66

.

0812-018

Figure 27

Female M12 Connector with Screw Terminals

List of PTB330 Serial Commands

The bold text in the brackets is the default setting. To issue a command,
type it on your terminal application and press the Enter key.

Measurement commands

INTV

[0...255 s/min/h/d] Shows or sets the continuous output interval (for RUN mode).

R

Changes the serial mode to RUN and starts outputting
measurement results according to the FORM string (with
interval defined by INTV).

S

Changes the serial mode to STOP and stops outputting
measurement results.

SEND

Shows the measurement results according to the configured
form.

Measurement setting commands

AVRG

[1...600 s]

Set the barometer measurement averaging time (in seconds)

DPMAX

Shows or sets the maximum allowed pressure difference
between barometer modules.

TQFE

[-80...+200 °C]

Shows or sets the temperature for QFE corrected pressure.

HHCP

[-30...+30 m]

Shows or sets the altitude for height corrected pressure.

HQFE

[-30...+30 m]

Shows or sets the altitude for the QFE corrected pressure.

HQNH

[-30...+3000 m]

Shows or sets the altitude for the QNH corrected pressure.