RADWAG MYA 5.3Y.F1 Microbalance User Manual

Page 174

Advertising
background image

- 174 -

Accessible responses:

PROFILE OK CR LF

- command understood, new PROFILE activated

LOGIN ERRROR CR LF - command understood, an error in PROFILE name,

log in failed

ES CR LF

- command not recognized (error in format)

The result release

Format: SS CR LF

Accessible responses:

SS_OK CR LF - command understood, realization in progress

The command simulates pressing PRINT button on an overlay, according to

settings selected for result release.

Internal adjustment/calibration

Format: IC CR LF

Accessible responses:

IC_A CR LF

IC_D CR LF

- command understood, realization in progress

- calibration completed

IC_A CR LF

IC_E CR LF

- command understood, realization in progress

- time limit of result stabilization exceeded

IC_I CR LF

- command understood but not accessible at this

moment

Lock automatic internal adjustment

Format: IC1 CR LF

Accessible responses:

IC1_I CR LF

- command understood but not accessible at this

moment

IC1_E CR LF

- unavailable operation, e.g. verified balance

IC1_OK CR LF

- command completed

For verified balances the automatic internal adjustment cannot be locked.

For non-verified balances the command locks internal adjustment, it may be

unlocked either by IC0 command or by switching the balance off.

The command does not influence settings responsible for adjustment process

beginning.

Unlock automatic internal adjustment

Format: IC0 CR LF

Accessible responses:

IC0_I CR LF

- command understood but not accessible at this

moment

IC0_OK CR LF

- command completed