Commissioning – IKA RV 10 digital FLEX User Manual

Page 30

Advertising
background image

30

The unit is ready for service when the
mains plug has been plugged in.

On/Off switch (Fig. 17)

Switch on the device on the right side of the device.

Device functions activated.

Function keys (see page 4)
• “Power” key (1)

Press the "Power" key once

“Stand-by” mode

Press the "Power" key again

Device functions reactivated

“Software version” display changes to "Parameter” display.

Note: The device must first be switched on at the on/off switch on the
right side of the device.

Note: The device is not dead or deenergised until the on/off switch has
been switched to off (right side of device)!

• Lift position key " " (2)

When you press the “ ”key, the lift goes up.

When you release the “ ” key, the lift stops and remains in that

position.

• Lift position key " " (3)

When you press the “ “key, the lift goes down.

When you release the “ “ key, the lift stops and remains in that

position.

• Rotating knob (4)

Setting the speed

Turn the rotating knob

Setting the speed

Speed setting accuracy: + 5 rpm,
Speed range 20-270 rpm
Note: When you select the speed > 100 rpm, smooth start is automati-
cally activated.

Start / Stop Rotation drive

Press the rotating knob

otation drive starts

Press the rotating knob again

Rotation drive stops

• Timer key (5)

Press the “Timer” key.

The timer block appears on the display, the “TIMER” indicator flashes.

Set the target value “Timer” on the knob by turning (1 to 199 minutes).

Save the “Timer” target value by pressing the "Timer" key again.

Starting the “Timer” function:

- Automatic when rotation drive is running
- Start the rotation drive.

Note: Once the timer has run out, there is a beep, the rotation finishes
and the lift goes up.

Commissioning

The previously saved value appears on the display.

Deactivate the timer by setting the target value to “0”.

• Int key (6)

Interval cycle for right-left running (change of direction) of the evapora-
tor piston in seconds.

Press the “Int” key.

The interval block appears on the display, the “INT” indicator flashes.

Set the target value “Int” on the knob by turning (1 to 60 minutes).

Starting the “Int” function:

- Automatic when rotation drive is running
- Start the rotation drive

Note: In interval mode the maximum speed is limited to 200 rpm.

Save the “Int” target value by pressing the "Int" key again.

Deactivate the interval by setting the target value to “0”.

Display (see page 4)

• Display "REMOTE" (10)

(only RV 10 digital)
Remote control via PC in connection with software “labworldsoft” is active.

• Display "888 rpm" (20)

Current target and actual speed indicator in revolutions per minute [rpm].
When the device is switched on the most recently selected target value
"speed" appears.

• Display (30)

Rotation drive active.

• Display "TIMER" (40)

Time lapse control active.
Indication of remaining distillation time in minutes [min].

Note: With the software "labworldsoft", RV 10 digital can also switch off
the heating bath once the timer has run out.

• Display "INT" (50)

Interval mode activated.

• Display "E01" (20)

--- not pictured ---
Error codes, see chapter entitled "Error Codes”.

Setting the lower end stop

Caution! Depending on the size of the piston, the setting angle of the

rotation drive and the position of the heating bath and lift, the eva-
porator piston can be in contact with the bottom of the heating bath.
Caution - glass may break!

Limit the lower lift position with the variable end stop.

Press the “ ” key until the lift reaches the desired position.

Note: The evaporator piston should be 2/3 immersed in the heating
bath.

To adjust the stop element (Q), press the centre button (R) on the front

of the lift (Fig. 18).

Move the stop (Q) to the desired position (Fig. 19)

Press the “ ” key until the lift has reached the upper end stop.

Note: The path is limited from 0 – 6 cm.

Fig. 17

Advertising