5.3

Thermal module

Based on the model, your epMotion can be equipped with one or more thermal modules.

5.3.1

Safety notes

5.3.2

Configuration

Abb. 5-7: Thermal module

Fig. 5-7:

Thermal module

Thermal modules temper liquids in a temperature range of 0 °C – 110 °C.

WARNING! Danger due to flammable or explosive liquids.

 Do not use explosive substances.
 Do not heat any highly flammable substances.
 Heat highly flammable substances in small quantities below boiling point only.
 Do not exceed the boiling point of solutions.

CAUTION! Burns from hot metal surfaces.
The metal surface heats up during operation. The software displays the current temperature of
the metal surface.

 Wait until the metal surface has cooled down completely.

CAUTION! Burns from hot labware.
Labware will be tempered during operation.

 Do not touch temperature-controlled labware.
 Wait until the labware has cooled down completely.

NOTICE! Material damage due to non-heat-resistant labware.
Samples and labware will become damaged if you do not use heat-resistant labware.

 Observe the labware specifications.
 At set temperatures in excess of 100 °C, only use labware made of metal, polypropylene

(PP) or polycarbonate (PC).