5.3.3 Home

The tools moves axially out of the working radius.
The function is intended for changing capillaries, or for quick removal of capillaries
out of the working radius.

HOME: Press (and hold down as necessary) to move out
as far as motors' end position.
The red LED flashes during the movement, and is then lit
steadily when the end position is reached.
Press the key again to move the tool back to its starting
point. To return the tool manually the joystick can be
moved directly. This terminates the Home function.

Clean moves the capillary out of the medium. In the
process, cell particles adhering to the medium boundary
layer are removed from the capillary.
When Clean is active:
CLEAN: Tap to move out a pre-set distance.
The red LED flashes during the movement and at the end
point.
Press the key again to move the tool back precisely to its
starting position, or move the tool manually.

If the parameter BackMan in the "Home" menu is set to
"off", it is then not possible to terminate the Home or
Clean function by moving the joystick
(see figure under notice text).
Interruption of the function is possible only by a
Reset ("Standby" button).

5.3.4 Axial

Activate or disable axial movement
Rotating the joystick causes the tool to
move axially up or down. The active
function is shown by the lit LED and is
indicated on the display.

Clockwise rotation:

Movement
diagonally downwards

Counterclockwise rotation: Movement

diagonally upwards

When the Z limit is reached the only movement is in X direction.

–3200
300

–21455

To move back use HOME

or joystick

HOME

–3200
300

–21455

To move back use HOME

or joystick

CLEAN

–3200
300
–21455

To move back use HOME

Home

coarse

X:

–1200µm

Y:

300µm

Z: –21323µm

axial

User control and mode of operation

5 User control and mode of operation

5