JUMO 902721 Test Equipment for Temperature and Temperature Traceability / Pt100 Precision RTD Temperature Probe Data Sheet User Manual
Page 2
2011-03-31/00331822
Data sheet 90.2721
Page 2/6
JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address: 36035 Fulda, Germany
Phone:
+49 661 6003-0
Fax:
+49 661 6003-607
e-mail:
Internet:
www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax:
+44 1279 635262
e-mail:
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard
Canastota, NY 13032, USA
Phone:
315-697-JUMO
1-800-554-JUMO
Fax:
315-697-5867
e-mail:
Internet: www.jumo.us
Precision display unit:
Description:
These universally applicable, process-controlled indicating devices convince with highest
accuracy and are perfectly suitable for measuring tasks that require highest precision. Ap-
plication areas are in the fields of quality assurance, production, maintenance/repair and
service as well as in the air conditioning and environmental technology. All devices of the
P700 series are equipped with a USB interface allowing an online documentation of the
measured values. This interface can also be used for supplying the devices with power.
The measured data can be processed further using the optionally available Windows soft-
ware "DE-Graph“.
Product features:
- integrated calibration function for simple compensation of probe tolerances
- 1-stage, 2-stage or 3-stage calibration optionally possible
- USB interface
- Windows evaluation software optionally available (see accessories "DE-Graph")
- Large LCD display
- Storage of MAX, MIN, HOLD and average values
- Differential temperature display for 2-channel units, simultaneous display of both val-
ues
- Mains operation possible
- Pt100 input in 4-wire circuit, thermocouple inputs as per DIN EN 60584
- Measuring channels can be freely assigned
- °C/°F change-over
Versions:
902722/20 (P700)
Pt100, thermocouple type J, K, L, N, R, S, T
1-channel; resolution 0.1 °C
902722/25 (P705)
Pt100, thermocouple type J, K, L, N, R, S, T
2-channel; resolution 0.1 °C
902722/30 (P750)
Pt100, thermocouple type J, K, L, N, R, S, T,
1-channel; resolution 0.01 °C from -200 to +200 °C,
otherwise 0.1 °C
902722/35 (P755)
Pt100, thermocouple type J, K, L, N, R, S, T
2-channel; resolution 0.01 °C from -200 to +200 °C,
otherwise 0.1 °C
Measuring ranges:
-200 to +850 °C (Pt100), thermocouples as per DIN EN 60584
Accuracy:
Version 902722/20 ... 25:
for Pt100
±
0.1 °C from -100 to +200 °C,
residual range 0.1 % of the measured value
for thermocouple type: R, S
± 1.0 °C + 0.1 % of the measured value
for thermocouple type: K, J, L, N, T
± 0.2 °C from 0 to 200 °C, ± 0.5 °C up to 1000 °C,
residual range ± 1.0 °C of the measured value
Version 902722/30 ... 35:
for Pt100
±
0.03 °C from -50 to +199.99 °C,
±
0.05 °C from -200 to -50.01 °C,
residual range 0.05 % of the measured value
for thermocouple type: R, S
± 1.0 °C + 0.1 % of the measured value
for thermocouple type: K, J, L, N, T
± 0.2 °C from 0 to 200 °C, ± 1.0 °C
residual range
Display:
2-line LCD display
Case:
ABS plastics, dimensions 200 mm x 93 mm x 44 mm (L x W x H)
Admissible operating temperature:
0 to 40°C
Weight:
approx. 350 g