3.6 Parameters, key <PARAM>

756/831 KF Coulometer, Instructions for Use

30


generator I: 400 mA






Oven: no

cond.










activate pulse: OFF

cond.

Current at generator electrode (100, 200, 400 mA, auto)

see also page 33.

auto

means that the current is automatically adapted to

the conductivity of the reagent and that in the region of

the endpoint the current will be controlled at smaller

values.

Connected oven (COM1, COM2, no)

COM of the Coulometer to which the oven is con-

nected.

If an oven is connected via RS232 an inquiry will be

made for the oven results and these will be inserted

into the result report of the Coulometer. The report out-

put on the oven must be switched OFF.

Set

no

if no oven has been connected or if you have not

connected the oven to Coulometer the via RS232 inter-

face.

Pulse output on I/O line L6 (L6, pin 1) of the remote

socket (first, all, cond., OFF)

see page 132.