3.6 Parameters, key <PARAM>
756/831 KF Coulometer, Instructions for Use
30
generator I: 400 mA
Oven: no
cond.
activate pulse: OFF
cond.
Current at generator electrode (100, 200, 400 mA, auto)
see also page 33.
auto
means that the current is automatically adapted to
the conductivity of the reagent and that in the region of
the endpoint the current will be controlled at smaller
values.
Connected oven (COM1, COM2, no)
COM of the Coulometer to which the oven is con-
nected.
If an oven is connected via RS232 an inquiry will be
made for the oven results and these will be inserted
into the result report of the Coulometer. The report out-
put on the oven must be switched OFF.
Set
no
if no oven has been connected or if you have not
connected the oven to Coulometer the via RS232 inter-
face.
Pulse output on I/O line L6 (L6, pin 1) of the remote
socket (first, all, cond., OFF)
see page 132.