Metrohm 940 Professional IC Vario ONE/Prep 1 User Manual

Page 39

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

940 Professional IC Vario ONE/Prep 1 (2.940.1110)

■■■■■■■■

31

Visually inspect whether all capillaries and their connections from
the high-pressure pump to the detector are leak-tight. If eluent is
leaking out anywhere, tighten the corresponding pressure screw
further, or loosen the pressure screw, check the end of the capil-
lary and shorten it using the capillary cutter if necessary and
retighten the pressure screw.

4 Conditioning the system

Rinse the system with eluent until the required stability of the base-
line is attained (normally 30 minutes).

The instrument is now ready for measuring samples.

Advertising