5 general information on separating columns, 6 selection of the sample loop – Metrohm 790 Personal IC User Manual
Page 43
2 Installation
790 Personal IC
34
2.7.5
General information on separating columns
New IC separating columns are normally filled with solution and
sealed at both ends. Before the column is installed in the system, it
must be ensured that this solution is freely miscible with the eluent
used (check manufacturer's specifications).
The IC separating columns and precolumns currently available from
Metrohm are listed in section 6.3.2. A test chromatogram and an infor-
mation leaflet is provided with each column. You will find additional in-
formation concerning these columns in the 8.732.2003 Metrohm Mono-
graph «Ion chromatography» and in special «Application Bulletins»,
which are available on request free of charge from your local Metrohm
agency.
When you install the column, always ensure that this is inserted
correctly in accordance with the flow direction shown (arrow must
point upwards).
2.7.6
Selection of the sample loop
Selection of the sample loop depends on the separating column used.
Normally, the following sample loops are used:
Cation columns
10 µL
Anion columns with suppressor
20 µL
Anion columns without suppressor 100 µL
Depending on the instrument version, the following sample loops are
installed in the 790 Personal IC:
Version
Sample loop
Volume
2.790.0010 6.1825.230
(PEEK) 10
µL
2.790.0020 6.1825.210
(PEEK) 20
µL
If desired, the built-in sample loop can be replaced by one of the sam-
ple loops available as an option (see section 6.3.1).