2.7 Precolumns and separating columns

790 Personal IC

35

2.7.7

Installation of the separating column without suppressor

With the 2.790.0010 Personal IC without suppressor module, the IC

separating column is installed as follows (see Fig. 13):

1 Connect column to injector

• Remove end caps from column 63

63

63

63.

without precolumn:

Screw inlet end of separating column 63

63

63

63 (note flow direction)

to column connection capillary 20

20

20

20 mounted on the injector.

with precolumn in cartridge head:

Install separating column 63

63

63

63 (note flow direction) in the car-

tridge head (see Fig. 11) as described in section 2.7.2.

with precolumn in cartridge holder:

Screw inlet end of separating column 63

63

63

63 (note flow direction)

to the precolumn fixed in the cartridge holder as described in

section 2.7.3 (see Fig. 12).

with IC anion precolumn SUPERSEP:

Screw inlet end of separating column 63

63

63

63 (note flow direction)

to the precolumn installed as described in section 2.7.4.

2 Rinse column

• Place a beaker beneath the column outlet.
• Open software window for manual system control.
• If necessary, modify

Flow rate

to the value suited for the

inserted separating column and click on

<Send to unit>

to

send this value to the 790 Personal IC.

• Switch on high-pressure pump (

IC pump

) by clicking

<On>

and

rinse column with eluent for ca. 10 min.

• Switch off high-pressure pump by clicking

<Off>

.

3 Connect column to detector block

• Screw outlet end of separating column 63

63

63

63 to the inlet capillary

37

37

37

37 permanently mounted on the detector block 38

38

38

38.

4 Fix column

• Insert one or two column holders 70

70

70

70 (6.2027.030, 6.2027.040

or 6.2027.050) in the mounting rails 19

19

19

19 and fasten separating

column 63

63

63

63 in the column holder 70

70

70

70.